Scopus:
Cochlear Synaptopathy Evaluation With Electrocochleography in Patients With Hearing Difficulty in Noise Despite Normal Hearing Levels

dc.contributor.authorYaşar, M.
dc.contributor.authorÖner, F.
dc.contributor.authorAtalay, F.
dc.contributor.authorAnbar, S.S.
dc.date.accessioned2024-12-12T09:16:32Z
dc.date.available2024-12-12T09:16:32Z
dc.date.issued2025
dc.description.abstractObjective: This study examined patients with normal hearing thresholds who had trouble understanding speech in noise. We used electrocochleography (ECochG) to detect and compare SP/AP amplitude area ratios, a potential indicator of cochlear synaptopathy, and investigate speech perception disorder in noise. Methods: The study included 68 people aged between 18 and 65 years, 35 patients and 33 healthy volunteers, who applied to the otorhinolaryngology clinic between November 2023 and March 2024 with a 2-month history of difficulty understanding speech in noisy environments. Everyone was given a tiptrode electrode ECochG test, and the results were compared between groups. An ECochG test was recorded with tiptrode electrodes and was performed on all participants, and the results were compared between groups. Results: In the ECochG test, the summation potential/action potential (SP/AP) amplitude and area ratios of patients who had difficulty understanding speech in a noisy environment were statistically higher than those of the control group. Conclusion: ECochG testing may provide additional evidence to evaluate auditory nerve pathways.
dc.identifier10.1111/coa.14235
dc.identifier.doi10.1111/coa.14235
dc.identifier.endpage81
dc.identifier.issn17494478
dc.identifier.issue1
dc.identifier.scopus2-s2.0-85205284804
dc.identifier.startpage75
dc.identifier.urihttps://hdl.handle.net/20.500.12597/33855
dc.identifier.volume50
dc.language.isoen
dc.publisherJohn Wiley and Sons Inc
dc.relation.ispartofClinical Otolaryngology
dc.relation.ispartofseriesClinical Otolaryngology
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectAction potential, cochlear synaptopathy, electrocochleography, summation potential
dc.titleCochlear Synaptopathy Evaluation With Electrocochleography in Patients With Hearing Difficulty in Noise Despite Normal Hearing Levels
dc.typearticle
dspace.entity.typeScopus
oaire.citation.issue1
oaire.citation.volume50
person.affiliation.nameKastamonu University
person.affiliation.nameKastamonu University
person.affiliation.nameKastamonu University
person.affiliation.nameKastamonu Training and Research Hospital
person.identifier.orcid0000-0003-3300-4430
person.identifier.orcid0000-0001-6195-3110
person.identifier.orcid0000-0002-0344-1982
person.identifier.orcid0000-0001-5671-7797
person.identifier.scopus-author-id59348789100
person.identifier.scopus-author-id57214728806
person.identifier.scopus-author-id57212937521
person.identifier.scopus-author-id59220226900

Files