Efficiency calibration of an HPGe detector in the 0.1–2.5 MeV energy range for Am–Be neutron source-based PGAA applications
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Journal of Radioanalytical and Nuclear Chemistry, Vol. 273, No.2 (2007) 443–450
DOI
Abstract
A 740 GBq 241Am–Be neutron source based prompt gamma-ray activation analysis (PGAA) setup in combination with a typical coaxial n-type HPGe detector (REGe) system was used to analyze light elements like H, B, C, N, etc. The absolute full energy peak (FEP) efficiencies of the shielded REGe detector for irradiation and counting geometries and for sources with different sizes (point, ampoule and cylindrical) were measured in the 0.1–2.5 MeV energy range by utilizing calibrated sources (point, liquid and solid). 4th order polynomials were fitted to the experimental data. Efficiencies in far irradiation and counting geometries are compared.
Description
Keywords
Citation
Collections
Endorsement
Review
Supplemented By
Referenced By
Google Scholar
Scholar'da Ara ↗ Bu yayında DOI yok — Altmetric/Dimensions/PlumX/BIP! rozetleri DOI gerektirir.
