Publication: The effect of the housing material on the NaI(Tl) detector response function
Loading...
Date
Organizational Units
Advisor
Editor
Journal Title
Journal ISSN
Volume Title
Type:
Publisher
Springer Science and Business Media LLC
DOI
10.1007/s10967-012-1716-z
Abstract
In this study, a Monte Carlo program has been developed to simulate the response function of the NaI(Tl) detector with all features for 60Co and 137Cs and to investigate the effects of detector housing material on response function. The pulse height spectra in a 2″ × 2″ NaI(Tl) detector due to these gamma ray sources have been measured. Comparison of the experimentally obtained and simulated spectra shows that there is good agreement between both spectra. Energy distributions of gamma photons that generate the backscattering peak in the response function were obtained and the contribution of single, double and multiple Compton scattering events to these distributions was investigated.
Description
Journal or Series
Journal of Radioanalytical and Nuclear Chemistry
ISSN
0236-5731
ISBN
Keywords
Citation
Collections
Endorsement
Review
Supplemented By
Referenced By
Rights and licensing
info:eu-repo/semantics/closedAccess